Download: User-Notice-Aquarion-Gravity-Fed-Eyewash-Assembly-Instruction
Encon Safety Products has identified a potential interference issue with the orientation of the Connector 2 components in some Encon Aquarion® Gravity Fed Eyewash systems. If you use or distribute the Encon Aquarion® Gravity Fed Eyewash station, please read this full user notice for instructions on how to determine if your unit is affected and the recommendations for solving a potential interference issue between the supply hose connector and the nozzle.
Root Cause and Corrective Analysis:
Our team has systematically analyzed the root cause of the interference condition, successfully identifying the contributing factor. The analysis underscores a direct relationship between the interference and the orientation of the Connector 2 components and the eyewash nozzles.
Verification and Corrective Steps:
At this time, to ensure the safety and operational functionality of your Aquarion® Gravity Fed Eyewash unit, please verify that the Connector 2 components are installed into the eyewash heads as pictured below. Examine the images and note the notch on the wing of the Connector 2 for the optimal orientation.
Aquarion Gravity Fed Eyewash System (Optimal Connector Orientation):
Optimal Connector 2 Orientation:
Interfered Connector 2 Orientation:
Recommended Action for Affected Units:
If you determine your Encon Aquarion® Gravity Fed Eyewash to be affected by this potential interference, we strongly advise the following steps:
- Disconnect the hose assembly from the eyewash head.
- Remove Connector 2 from the hose assembly.
- Rotate Connector 2 by 180 degrees.
- Reattach Connector 2 to the hose assembly.
- Reconnect the hose assembly to the eyewash head.
This corrective action is intended to rectify the identified interference condition.
Notice to Distributors:
We request all distributors to promptly communicate this user notice to their customers who have purchased the Encon Aquarion® Gravity Fed Eyewash. Distributors play a crucial role in making end-users aware of this potential interference.
For further assistance or inquiries, please contact our Customer Service team at customerservice@enconsafety.com or call 1-800-283-6266.
We sincerely appreciate your cooperation in ensuring the continued safety of Encon Aquarion® Gravity Fed Eyewash users.